PANALYTICAL XPERT3 POWDER XRD

Restart procedure for PXRD (i.e. x-ray power shows “0 kV, 0 mA”)

FOR SCI ERROR, PLEASE
CLICK HERE!

If the doors are not properly closed, either too gently or too harshly, the microswitches in the door safety system can have trouble making contact and this will shut off the x-rays (as a safety feature). If this occurs while using, please try the following procedure:

  1. Symptom: X-rays go to zero

  2. Turn the HV key off

  3. Cycle the doors (open/close)

  4. Turn the key to on

  5. The light bar near the key and the top light should light up in a few seconds. If it does not, turn the key off and repeat from the door cycling step

  6. When the lights go on, the x-rays will automatically go to 30 kV, 10 mA.

  7. After connecting, the x-rays can be set to 45 kV and 40 mA (or 10 mA if going to standby mode).

  8. If a user cannot get the x-rays to restart, contact staff via Badger.